Instrument for contact temperature measurement

Manufacturer and type:National Instruments NI 4350
Purpose:Contact temperature measurement
Kratki opis metode:The temperature measurement method in this device is based on the creation of thermo voltage at the ends of the thermocouple as a measuring sensor. Thermocouples connect to a multi-channel device (AD converter) that is connected to a notebook computer that memorizes and processes the measured values.
Tehničke značajke:Thermocouples connect to a multi-channel device for precise temperature measurement of NATIONAL INSTRUMENTS. The device consists of a multichannel analog input port TC-2190 and a voltage meter type NI 4350. By measuring of temperature, following type of thermocouples can be connected: J, K, N, E, T, R, S and B which allow us to measure the temperature range of -100 °C to + 1750 °C with a measurement accuracy within 0.1%. The device has the ability to read up to 60 values per second. The measurement results are stored and processed on the connected laptop.
Standard:DIN 43760
Tip i priprava uzorka:No sample preparation
Laboratory:Laboratory for heat engineering and mechanical engineering
Operater:
Previous apparatus Next apparatus
« »